Share
print this page decrease the text size increase the text size

Riata and Riata ST Silicone cardiac leads - increased wear

Safety advisory

17 May 2012

Name Model number
Riata Silicone 1560 1561 1562 1570 1571 1572 1580 1581 1582 1590 1591 1592
Riata ST Silicone 7000 7001 7002 7010 7011 7040 7041 7042

The TGA advises patients and health care professionals of potential problems with the above models of Riata and Riata ST Silicone cardiac leads (manufactured by St Jude Medical) for implantable cardiac cardioverter-defibrillators (ICDs). Please note that this advisory does not apply to Riata ST Optim leads.

In view of the relatively low failure rate and the risks associated with surgery to replace the implanted leads, routine removal of these leads is not recommended in the absence of evidence of a fault. However, all patients with an ICD using Riata or Riata ST Silicone leads should make an appointment with their cardiologist in response to this alert.

The supply of these particular leads was discontinued in Australia in December 2010. These devices have been cancelled from the Australian Register of Therapeutic Goods (ARTG). Any Riata leads implanted after this date would be of another type and are not affected by this advisory.

TGA is continuing to investigate the reports associated with the failure of the Riata and Riata ST Silicone leads.

Potential problems

There are three potential ways that ICDs connected to these leads may operate incorrectly:

  1. the ICD may deliver additional electrical charges when they are not required by the patient;
  2. the ICD may not operate when required; or
  3. may not provide effective therapy.

These potential problems can be caused by the leads becoming exposed by two different wear patterns:

  1. Externalisation (extrusion) of the conductor outside the implanted lead; or
  2. Wear (abrasion) to the insulation surrounding the lead.

Information for consumers

If you have an implanted ICD:

If your ICD has Riata or Riata ST Silicone leads connected:

Patients with any ICD should carry their identification card with them at all times. If you have any concerns, regardless of the type of defibrillator leads you have, contact a health professional, preferably the cardiologist who implanted the device.

Information for interventional cardiologists

St Jude Medical (the manufacturer), has advised that:

St Jude Medical has written to implanting surgeons/cardiologists and hospitals where the implantations were undertaken. A copy of St. Jude Medical's hazard alert communication, which contains additional information, follows.

Information for all health professionals

Patients with an ICD using Riata or Riata ST Silicone leads should be referred to a cardiologist, preferably the implanting cardiologist. Note that this advisory does not apply to Riata ST Optim leads.

Technical information

There are several factors that can contribute to lead wear and abrasion in implanted pacing and defibrillation systems, including physiological stresses placed on the lead due to patient anatomy, implant orientation, and mechanical stresses applied from concomitant devices in the body. The main causes of insulation problems are listed below:

Reported rates of these problems

The manufacturer, St Jude Medical, indicates that a total of 3962 Riata and Riata ST leads were supplied in Australia over the last 9 years. Over 227,000 Riata and Riata ST leads have been supplied world-wide during that time.

Rates of lead failure are likely to be under estimated because not all cases are reported to either the manufacturer or to TGA. However, St Jude Medical has provided estimates of their confirmed incident rates for Australia and world-wide. St Jude Medical reports that the Australian reported incident rate up to 29 January 2012 is estimated to be 1.2% for all cause abrasion with approximately 18% of those exhibiting externalised leads. St Jude Medical has estimated that world-wide (including Australia) reported incident rate up to 30 November 2011 is 0.63% for all cause abrasion, with approximately 15% of those exhibiting externalised leads.

There is some debate in the medical literature1,2, regarding the rate of these problems, however the Riata and Riata ST Silicone leads have been removed from the ARTG and have not been available in Australia since December 2010. Patients are advised to speak to their cardiologist regarding any concerns they may have.

Reference

  1. Hauser, R (et al); Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads; (accepted manuscript 4 March 2012; Heart Rhythm)
  2. Carlson, M: ICD leads and postmarketing surveillance (letter to the editor): NEJM; 8 March 2012

Reporting problems to the TGA

Consumers and health professionals are encouraged to report problems with medical devices. Please refer to the TGA Internet site for more information on the TGA Incident Reporting and Investigation Scheme (IRIS).

The TGA cannot give personal advice about an individual's medical condition. You are strongly encouraged to talk with a health professional if you are concerned about a possible adverse event associated with a medical device.

Top of page

Content last updated: Thursday, 17 May 2012

Web page last updated: Thursday, 17 May 2012

URL: http://www.tga.gov.au/safety/alerts-device-riata-cardiac-leads-120517.htm

Site navigation

Safety information

Consumers

Health professionals

Industry

About the TGA

News room