Providing patient information leaflets and implant cards for medical devices
Guidance on how to provide the required patient information materials with all implantable and active implantable medical devices.
Recently published
This page was published on [date_placeholder].
Recently updated
This page was updated on [date_placeholder]. See page history for details.
Purpose
The purpose of this guidance is to help manufacturers and sponsors understand how the TGA interprets the legislative requirements for patient information materials (patient implant cards and patient information leaflets) and thus understand how they can comply with the legislative requirements.
This is a guide only; manufacturers and sponsors should familiarise themselves with the legislative and regulatory requirements and, if necessary, seek professional advice. It is the responsibility of sponsors and manufacturers to understand and comply with these requirements.
This guidance will continue to be reviewed and revised, where necessary.
This guidance provides an overview of:
- the different types of patient information materials (patient information leaflets and patient implant cards)
- when patient information must be supplied
- how to meet the mandatory requirements for patient information
- best practice requirements for patient information and
- what to do if your patient information materials are not compliant.
This guidance refers to requirements set out in clause 13A of Schedule 1 of the Therapeutic Goods (Medical Devices) Regulations 2002 (the MD Regulations).
Our Acronyms and glossary page may be useful to clarify the terms used in this document.
Legislation
Patient information materials
Purpose
Patient information materials consist of:
- patient information leaflets and
- patient implant cards.
Patient information materials assist patients to:
- understand the medical device being implanted, both prior to and following surgery
- have informed consent conversations with their health professional and
- report any adverse events associated with their implanted medical device.
Manufacturer and sponsor responsibilities
The requirements for patient information materials are part of the essential principles in Part 2 of the MD Regulations. Sponsors must ensure that the kind of device included in the Australian Register of Therapeutic Goods (‘the Register’) under their name complies with the essential principles, including the requirements for patient information materials.
Manufacturers are responsible for creating the content of the leaflets and cards. The manufacturer is also responsible for maintaining the currency of this information in these materials and ensuring that they hold the evidence to support the information provided. Sponsors must ensure that they have available sufficient information to substantiate that the devices comply with the essential principles (including the patient information material requirements), or have procedures in place to obtain any relevant information from the manufacturer to substantiate that the materials comply with the requirements in the essential principles.
When patient information materials are required
On 1 December 2018, regulations1 commenced to require patient information materials to be supplied with:
- implantable medical devices and
- active implantable devices.
A graduated transition period applied as described in Attachment 1.
Implantable and active implantable devices are defined in the Therapeutic Goods (Medical Devices) Regulations 2002 as follows:
implantable medical device means a medical device (other than an active implantable medical device) that is intended by the manufacturer:
- to be, by surgical intervention, wholly introduced into the body of a human being, and to remain in place after the procedure; or
- to replace, by surgical intervention, an epithelial surface, or the surface of an eye, of a human being, and to remain in place after the procedure; or
- to be, by surgical intervention, partially introduced into the body of a human being, and to remain in place for at least 30 days after the procedure.
active implantable medical device or AIMD means an active medical device, other than an implantable medical device, that is intended by the manufacturer:
- either:
- to be, by surgical or medical intervention, introduced wholly, or partially, into the body of a human being; or
- to be, by medical intervention, introduced into a natural orifice in the body of a human being; and
- to remain in place after the procedure.
Note
The definition of implantable medical device includes devices that are wholly or partially absorbed by the body.
Implantable custom made devices and patient matched devices are required to meet the requirements for patient information materials.
Implantable medical devices exempted from this requirement
The following implantable devices, and articles similar to these, are exempted from the obligation to provide the patient information materials:
- sutures
- staples
- dental fillings
- dental braces
- tooth crowns
- screws
- wedges
- plates
- wires
- pins
- clips
- connectors
- a medical device that is intended by the manufacturer to be wholly, or mostly, absorbed by a patient’s body within 6 months of being implanted; and is for use as a filler, haemostasis, tissue approximation, fixation of other medical devices within tissue.
Attachment 2 provides further details of how the TGA interprets the above.
Note
While these devices are exempted from the statutory requirement to provide patient information materials, the TGA strongly encourages manufacturers and sponsors to provide compliant patient information materials for these devices as a matter of best practice. Doing so will mean that patients and health practitioners using these devices will benefit from information about the devices in the same way as patients and health practitioners do for non-exempted implantable devices.
Patient information materials that are non-compliant with the Essential Principles
From 1 December 2021, all implantable medical devices are required to have patient information materials available in the form of both Patient Information Cards (PIC) and Patient Information Leaflets (PIL). This date signifies the end of the transition period noted in the above-mentioned Regulation changes (see transition period described in Attachment 1).
For sponsors of medical devices currently in the ARTG, who will not have compliant PICs/PILs in place by 1 December 2021, the TGA will consider approving a consent to import, supply, or export a medical device that does not fully comply with the regulatory requirements. Authorised representatives of a sponsor can apply for a consent to import, supply, or export a medical device that does not comply with the Essential Principles. To ensure continuous supply of an affected device, the consent will need to be in place prior to 1 December 2021.
There are criminal offences under section 41MA and civil penalties under section 41MAA of the Therapeutic Goods Act 1989, for persons who import, supply, or export medical devices that do not meet the Essential Principles for safety and performance, unless consent has been granted by the Secretary of the Department of Health.
Consent can also be sought for devices which have an Application for Inclusion in the ARTG. The consent must be in place prior to the device being approved and included in the ARTG.
Electronic Patient Information Cards and Leaflets
On 29 October 2021, amendments to the Therapeutic Goods (Medical Devices) Regulations 2002 came into effect to allow patient information materials for implantable and active implantable devices to be supplied in more flexible (principally electronic) formats. The Regulation changes mean that patient information cards and patient information leaflets can be supplied electronically rather than in hard copy format, as long as they contain all required information and are made available in a way that is readily accessible by the patient concerned. Where this is the case, devices supplied with electronic PICs and/or PILs in a way that is readily accessible by the patient concerned will be considered compliant and consent will not be required for these devices.
Consent application forms
The TGA has modernised the consent application process, moving from a paper form to an online form hosted in the TGA Business Services (TBS) portal.
The link to the consent application form is also available on our website at Essential Principles: consent for non-compliance.
Application Fees
Reduced consent application fees for devices with non-compliant patient information materials.
On 29 October 2021, amendments to the Therapeutic Goods (Medical Devices) Regulations 2002 came into effect to introduce a fee concession for sponsors of implantable devices seeking consent to import, supply, or export their devices where they do not have compliant patient information materials. The application fee has been reduced to a flat fee of $30 for each ARTG entry and Application for Inclusion where the application is made solely in relation to non-compliance with EP 13A.2 and/or 13A.3.
The TGA will apply this fee concession retrospectively, refunding the difference in fees to eligible sponsors who have lodged a consent to supply application solely in relation to non- compliance with EP 13A.2 and/or 13A.3 on or after 1 January 2021. Sponsors that believe they are eligible for a fee refund should contact the TGA at mdconsent@health.gov.au identifying the relevant consent application. If the refund request is validated, it will be forwarded to the Product Billing and Industry Assistance team who will contact you regarding the refund.
NOTE: For consent applications relating to other Essential Principles (in addition to, or different to, EP 13A.2 and/or 13A.3), the normal processing fees of $500 for the first and $100 for each subsequent ARTG entry/Application for Inclusion applies.
How do I calculate my fees?
For consent applications relating to non-compliant patient information materials, the fee is $30 per ARTG entry / Application for Inclusion. For example, if your application for consent pertains to 45 ARTG entries, the fees are calculated as $30 x 45 = $1,350.
How to pay
There are two ways to pay the application fee for the consent to supply application.
- IMMEDIATE PAYMENT – You can pay the applicable application fee for your consent application immediately after completing and submitting the final form in your application. Your application and payment will be linked during processing using the ARTG entry or Application for Inclusion provided in the payment details.
- PAYMENT AGAINST INVOICE - If you require the TGA to raise an invoice for payment, simply complete and submit your application for consent, and the TGA will raise and send the submitter an invoice for the processing fees.
NOTE: Applications for consent will not be processed until all applicable fees have been paid in full.
Note about patient implant cards:
In cases where a physical patient implant card is provided, containing the manufacturer’s information, with the device and a sticker, containing the device information, is also provided with the device, but the sticker is required to be adhered to the card at the point of care, this is considered compliant with the legislated requirements.
Implementation Plans
Applications for consent will require the submission of an Implementation Plan which details how and when the device(s) will comply with the regulatory requirements for both the PIC and the PIL, and what interim arrangements are in place to enable patients and healthcare providers to access the relevant patient information materials. It is anticipated that a single Implementation Plan may apply to multiple ARTG entries/Applications for Inclusion. As part of the consent application process, sponsors must provide copies of the interim non- compliant patient information, if applicable, to support the information provided in the Implementation Plan.
For example, if the consent relates to non-compliant patient implant cards due to missing device information, it would be expected that the Implementation Plan would include:
- what information is missing from the card (i.e. the card is missing the device model and batch number)
- how the missing information is going to be provided to the patient or healthcare facility and when will this be provided (e.g., the information is on stickers with the device, but as there are insufficient stickers to adhere to the card, this information will need to be hand written on the card)
- how the non-compliant card will be provided to the patient or healthcare facility and when will this be provided (e.g., a template card will be provided in bulk prior to device being supplied or with the device. Extra cards can be requested from the sponsor or downloaded from the manufacturer’s web site)
- what the expectation is of the healthcare facility (if any) to facilitate the convergence of the missing information with the non-compliant card (e.g., the healthcare facility will need to write the information on the template card)
- will there be education sessions provided to the healthcare facility or patient, and if so, when will these be provided (e.g., the sponsor will provide support to the healthcare facilities when products are ordered or supplied)
- when can the healthcare facility or patient expect to receive a compliant patient implant card and how will this change be relayed (e.g., the compliant cards will be introduced for products manufactured from 20 May 2022. Healthcare facilities will be advised by email when the new cards are being included with the device and they are no longer required to add stickers to template cards) and
- a copy of interim non-compliant card, if applicable.
Monitoring of Implementation Plans during the consent period.
If consent is approved, it is expected that sponsors will work towards supplying compliant patient information materials within the consent period as outlined in their Implementation Plan. The TGA will monitor sponsor Implementation Plans throughout the approved consent period to ensure that sponsors are following their plans. Sponsors are required to submit evidence to the TGA of compliant patient information materials at the end of the consent period. Instructions regarding the submission of documents for evidence of compliance at the end of the consent period are provided during the consent approval process.
Compliance with advertising legislation
Patient information leaflets and cards are not intended for advertorial or promotional purposes.
Manufacturers and sponsors can reduce the risk of their content being considered promotional (and therefore an advertisement) by:
- presenting only information in a factual and balanced manner
- not including information about different therapeutic options in a way that implies that the medical device implant is the best option
- providing a balanced overview of the therapeutic options and their place in recognised therapeutic regimes. This can be provided in supporting materials, but comparative statements (e.g. newer/ more effective/better tolerated/ more evidence to support use than XXX, etc.) should not be used and
- ensuring a leaflet that is non-promotional in content does not inadvertently (or unintentionally) become part of an advertisement if, for example, it is published on a sponsor website with promotional statements about the company’s superior manufacturing characteristics, etc. It can also become part of an advertisement if it is presented in a way that facilitates patients ‘shopping’ for a device that might address their disease, condition, ailment etc.
Check the Australian Regulatory Guidelines for Advertising Therapeutic Goods for more guidance about the characteristics of content that is likely to be considered promotional.
If you are concerned that a leaflet may be considered promotional or could be found to be used in a promotional way, you should refer to the Therapeutic Goods Advertising Code (the Advertising Code).
Leaflets and cards that are non-compliant with advertising legislation
Where the leaflets or cards appear not to comply with the advertising legislation, follow-up actions may be undertaken by the TGA.
For more information about complaints, go to the TGA’s Advertising hub.
Patient information leaflets
Purpose
The patient information leaflet should be one of many sources of information that inform a discussion on the decision regarding the implantation of a device.
It is considered best practice to make leaflets available to doctors and potential patients prior to surgery to assist patient-doctor discussions regarding:
- the type of medical device being considered and
- the type of medical condition the device is used for.
The leaflet may also be used to provide patients with:
- the name and manufacturer of the device
- information about what may happen after the surgery and
- information about possible adverse events and malfunctions.
Mandatory requirements for leaflets
A patient information leaflet must be written in English, and may also be provided in any other language. It may also include diagrams, drawings or symbols (e.g. MR status symbols).
If a patient information leaflet is supplied in hard copy, the leaflet must be written in English, and may also be provided in any other language
- have text that is legible and at least 1 millimetre high. ‘Text’ includes any:
- number
- letter
- symbol
- letter or number in a symbol.
You must:
- ensure the leaflet is available and readily accessible by the patient and ensure that the leaflet is written in a way that is readily understood by patients.
The leaflet must include:
- information identifying the device, or the kind of device
- the intended purpose of the device
- information explaining how to use the device safely and
- other information about the device that the manufacturer considers would be useful for patients.
- in particular, the leaflet must include the information listed in the table below (see Clause 13A.3 of the MD Regulations).
Item | Information to be included in leaflets |
---|---|
1 |
|
2 |
|
3 | Any special operating instructions for the use of the device. |
4 |
|
5 | Any residual risks that could arise due to any shortcomings of the protection measures adopted as mentioned in subclause 2(2)3 |
6 |
Example 1 The risk of electrical interference from electro surgical devices. Example 2 The risk of magnetic field interference from magnetic resonance imaging devices. |
7 |
|
8 |
|
9 |
|
For additional ways to make the leaflets user-friendly: please see Best Practice for Patient information leaflets and patient implant cards, at Attachment 3.
Complete information
It is expected that the patient is provided with full and complete information about their device, without the need to refer to further information. Statements such as “consult your doctor about possible side effects” and “please see the full list of precautions and contraindications in the instructions for use” are not appropriate, as the patient may not have access to these resources. Warnings about magnetic resonance (MR) conditions should be present, and provided in a way which is readily understood by patients. Incomplete information or references to alternative sources make it more difficult for the patient to access the required information and the TGA may consider that the information in the leaflet is not written in a way that is readily understood by patients (see Clause 13A.3(4), Schedule 1, part 2 of the MD Regulations).
Adverse events – urogynaecological meshes and breast implants
The table of mandatory information for patient information leaflets requires that any undesirable side effects that could be caused by use of the device must be included in the patient information leaflets (see Item 4b, Clause 13A.3 of the MD Regulations).
For urogynaecological meshes and breast implants, the TGA expects that, to comply with Item 4b, manufacturers must include certain known adverse events for these devices in the patient information leaflets. Known adverse events that have been derived from the extensive post-market reviews of these products are listed in Attachments 4 and 5. However, these lists are not intended to be exhaustive - manufacturers are obliged to review and update the lists if further undesirable side effects arise over time.
How and when to provide leaflets
You must provide the patient information leaflet to the patient concerned, however the MD Regulations do not prescribe the manner in which the leaflet must be provided.
The TGA expects sponsors and manufacturers to ensure the leaflet:
- can be readily accessed by consumers and healthcare professionals
- can be accessed free of charge
- is available as early as possible, so that medical practitioners and patients can use it to inform their discussions on the proposed course of treatment.
Electronic leaflets
Where leaflets are provided in hard copy, sponsors are strongly encouraged to provide electronic patient information leaflets to enable early access to information for healthcare professionals and patients.
When providing electronic leaflets, ensure:
- patients are made aware of how to access the electronic versions
- patients can easily navigate the manufacturer’s website and find the correct leaflet.
Sponsors and manufacturers are expected to keep sufficient information to establish that:
- electronic leaflets have been provided with the device
- the requirements of clause 13A.3 of the MD Regulations have been met.
Sponsors may be requested to provide this information to the TGA.
Note
Like hard copy leaflets, electronic patient information leaflets must contain the information required in clause 13A.3 of the MD Regulations. On 29 October 2021, amendments to the Therapeutic Goods (Medical Devices) Regulations 2002 came into effect to allow patient information materials for implantable and active implantable devices to be supplied in more flexible (principally electronic) formats.
Leaflets for a ‘kind of device’
It is permissible to have one patient information leaflet to cover multiple devices if they meet all of the following criteria. They:
- are manufactured by the same manufacturer
- have the same sponsor
- have the same device classification
- have the same device nomenclature system code6
- share the same intended purpose and
- share the same warnings, precautions, and user risks.
The leaflet should:
- clearly identify the devices intended to be covered by the leaflet and
- list the name and model of each device.
Date stamping and version control
For both hard copy and electronic patient information leaflets, you should:
- clearly state the date of release of the information
- have processes in place for version control and
- ensure earlier versions of the document (even those for products considered obsolete) remain accessible to the public.
Patient implant cards
Purpose
A patient implant card is a card intended to be provided to a patient following surgery when the patient has received:
- an implantable medical device or
- an active implantable medical device.
The purpose of patient implant cards is to ensure that patients are aware of the details of the device that they have been implanted with and that health practitioners can also identify particular devices. The cards should also better enable the traceability of the device and patient in order to more quickly and effectively alert patients and health practitioners to safety issues such as precautions or recalls.
Mandatory requirements for implant cards
A patient implant card (see clause 13A.4 of the MD Regulations):
- must be available and readily accessible by the patient
- must be written in English, and may also be provided in any other language
- may also include diagrams, drawings or symbols (e.g. MR status symbols)
If a patient implant card is supplied in hard copy, the card must:
have text that is legible and at least 1 millimetre high. ‘Text’ includes any:
- number
- letter
- symbol
- letter or number in a symbol.
You must include the following information on the card (see Clause 13A.2 of the MD Regulations):
- the name of the device
- the model of the device
- the batch code, lot number or serial number of the device and
- the manufacturer’s name, address and website address.
Other details you may wish to include
Although not required, you may also include the sponsor’s details if you wish.
There is also no requirement to include warnings on the patient implant card. You may decide to include some warnings, for the patient’s benefit, where it is appropriate (e.g. about possible interactions with other electronic equipment such as airport security scanners or magnetic resonance imaging (MRI) equipment for pacemakers or intra-ocular lenses).
There are additional things that you can do to make the cards user-friendly: please see Best Practice for Patient information leaflets and patient implant cards, at Attachment 3.
How to provide implant cards
The patient implant card should be provided to the patient concerned as soon as practical. This will allow health professionals and patients rapid access to the information.
You may wish to provide additional space on the card for healthcare professionals to insert:
- the name of the surgeon and
- the name of the hospital where the procedure was undertaken.
You may supply bar codes on stickers with a device as a means of identifying the device. This is acceptable, provided the stickers:
- are durable and
- contain the required information and are in the correct form: see Mandatory requirements for implant cards.
In cases where physical patient implant cards containing some of the information and stickers containing the remainder of the required information are both provided with the device, but the sticker is required to be adhered to the card at the point of care, this is considered compliant.
Electronic patient implant cards
In addition, or alternatively, to physical cards, you may also provide patient implant cards electronically. However, the card must include all required information (see Clause 13A.2 of the MD Regulations). Care must be taken to ensure that electronic patient implant cards are able to be easily accessed by the patient from the provided website and the correct information/card can be found.
Review by the TGA
Patient information leaflets and patient implant cards will be assessed when the TGA undertakes assessment or review of medical devices as part of its regulatory activities. This includes during:
- TGA’s conformity assessments
- application audits
- post-market reviews.
Sponsors must be able to obtain the required information from the manufacturers and provide it to the TGA if requested, in order to demonstrate compliance with the essential principles in the MD Regulations.
If the manufacturer holds a conformity assessment certificate issued by the TGA for implantable devices that are already supplied in Australia, applicants will be required to include patient information leaflets and patient implant cards as part of the ‘Information to be provided with medical devices’ that is routinely reviewed during an application for recertification of an existing conformity assessment certificate. Manufacturers may also be asked to submit patient implant leaflets or cards for review as part of any other regulatory activity.
Safety related changes should be managed in accordance with the Uniform Recall Procedure for Therapeutic Goods, to ensure appropriate notification is provided to affected consumers.
Reporting adverse events
Adverse events are unintended and sometimes harmful occurrences, associated with the use of a therapeutic good, and include incidents involving medical devices.
The patient information materials will be useful to patients and healthcare professionals lodging an adverse event report. Certain devices (e.g. urogynaecological meshes and breast implants) are subject to specific requirements about which known adverse events should be included in the patient information leaflets, so as to comply with the requirement to disclose undesirable side effects – see Attachments 4 and 5.
Sponsors must report adverse events to the TGA, but anyone can report a suspected adverse event.
Reports should include as many details as possible including:
- contact details for the reporter to assist the TGA in case follow up information is required;
- a description of the adverse event; and
- details of the medical device suspected of causing the adverse event. Go to Reporting adverse events on our website for more information on how to report an adverse event.
Attachment 1: Timetable for transition
From 1 December 2018, manufacturers and sponsors of all new implantable or active implantable medical devices (other than those exempted) have been required to make available to patients, patient information leaflets with the device.
A “new” device means one that:
- is in the Register because of an application made on or after 1 December 2018.
From 1 December 2018, manufacturers and sponsors of new urogynaecological mesh devices have been required to provide patient implant cards with their devices. New devices that are not urogynaecological mesh devices must be accompanied by patient implant cards from 1 December 2020.
A graduated transition period applies for existing medical devices. An “existing” device means one that:
- is in the Register because of an application made before 1 December 2018, regardless of the date the device was included in the Register (referred to as a pre-commencement entry in the MD Regulations).
From 1 December 2021, all implantable or active implantable devices will require patient implant cards and patient information leaflets, unless they are specifically exempted from these requirements. Some examples are included below the table.
Patient Information Leaflet (PIL) | Patient Implant Card (PIC) | |
---|---|---|
Urogynaecological mesh | ||
New devices | 1 Dec 2018 | 1 Dec 2018 |
Existing devices | 1 Dec 2019 | 1 Dec 2019 |
Surgical mesh | ||
New devices | 1 Dec 2018 | 1 Dec 2020 |
Existing devices | 1 Dec 2021 | 1 Dec 2021 |
Implantable medical devices (other than those exempted) | ||
New devices | 1 Dec 2018 | 1 Dec 2020 |
Existing devices | 1 Dec 2021 | 1 Dec 2021 |
Examples
- a device is included in the Register on 15 July 2018 – it is an “existing” device – a PIC and PIL must be provided with the device from 1 December 2021.
- An application to include a device is made on 20 October 2018 and the device is included in the Register on 20 December 2018 – it is an “existing” device – a PIC and PIL must be provided with the device from 1 December 2021.
- An application to include a device is made on 22 May 2020 – it is a “new” device and requires a PIC to be provided with the device from 1 December 2020 and a PIL immediately (i.e. from commencement of supply).
- An application to include a device is made on 12 December 2020 – it is a “new” device and requires a PIC and PIL to be provided with the device immediately (i.e. from commencement of supply).
Attachment 2: Implantable medical devices exempted from requirements
A number of implantable devices are exempted from the obligation to provide patient information materials (patient information leaflets and patient cards).
Clause 13A.1(1)(b) and 13A.1(2) of Schedule 1, part 2 of the Therapeutic Goods (Medical Devices) Regulations 2002 lists the exempt devices:
- suture, staple, dental filling, dental brace, tooth crown, screw, wedge, plate, wire, pin, clip, connector or similar article
- a medical device that is intended by the manufacturer to be wholly, or mostly, absorbed by a patient’s body within 6 months of being implanted; and is for use as a filler, haemostasis, tissue approximation, fixation of other medical devices within tissue, or a similar medical device.
This attachment explains how the TGA interprets the terms listed in the MD Regulations.
Notes
As a matter of best practice, the TGA strongly encourages manufacturers and sponsors to provide patient information materials compliant with the essential principles for ALL devices (including those listed in this attachment).
The information provided in this attachment does not include an exhaustive list of devices which may/may not be exempt.
Please contact us if you require clarification regarding a specific device.
Sutures
Sutures are considered to be any structure intended to hold together two opposing ends. For example:
- the two ends of a wound
- the rotator cuff to the humerus.
Sutures have other characteristics. They can be:
- monofilament/multifilament
- absorbable/non-absorbable
- natural/synthetic.
Sizes vary for both the suture thickness and the needle that is attached.
Sutures can be attached to other devices, such as in the case of a suture anchor, where a suture is attached to an anchoring device. The suture in this case still has the same purpose. It aims to bring together two or more structures together (depending on how many sutures are attached to the suture anchor). A suture anchor is not considered to be a suture.
Staples
Staples are considered to be a device identical in appearance to the household stationary ‘staple’. They serve to hold two ends together, for example two bones in the feet during fusion surgery. Staples can also be used instead of sutures to hold a wound together, or to fix mesh onto tissues such as in the case of a hernia.
They are usually of metallic composition.
Dental fillings
Dental fillings are considered to be any filling substance used to repair a tooth cavity.
Dental braces
Dental braces are considered to be a combination of brackets and wires, or other materials, used externally on the tooth to correct alignment. This includes both the traditional metallic braces and non-metallic braces, such as ‘Invisalign’.
Tooth crowns
Tooth crowns are considered to be any device that covers an existing tooth.
Screws
Screws are considered to be a monoblock device with a raised helical thread intended to fix two solid objects together. They have a slotted head, to allow for tightening using a driver.They can achieve this fixation in combination with a plate or rod.
Medical screws are often made from a metal such as stainless steel or titanium, however, recent advancements have seen the growth of biodegradable screws used in specific applications.
Exempt |
---|
Blocking screw |
Cap screw |
Set screw |
Wedges
Wedges are considered to be a device, with a constant/uniform thickness or tapering thickness, that is inserted between two structures to secure or separate them.
Plates
Plates are considered to be a flat or contoured device with screw holes that are used to provide reduction, stability and fixation.
They can have varying thicknesses and are usually composed of stainless steel or titanium. They can accommodate different types of screws (e.g. locking or compression screws). They can be malleable or rigid.
Wires
Wires are considered to be singular continuous pieces of metal (identical to the everyday wire) used to re-attach bone fragments or provide stability.
Pins
Pins are considered to be a straight piece of metal used to stabilise bones. They are commonly used to hold a reduction whilst the surgeon attempts to achieve more permanent and reliable fixation (pin is removed once fixation is achieved). Very rarely, they are used independently by surgeons for permanent fixation.
Clips
Clips are considered to be a device used to hold a part or thing together with another.
For example, an arterial clip is used to close off a small vessel (i.e. bring the walls of the vessel together).
Connectors
Connectors are considered to be devices which attach two or more different components to one another.
Similar articles
Exempt | Not Exempt |
---|---|
Anchor/suture Anchor/clamp Transverse cross links/cross connectors Cleat Button/cable plug ICD adaptor ICD extender ICD Splitter Dental abutment Fracture pins Cables Cerclage wires Lead anchoring sleeve Laminar hook Grip plate Occipito-cervical and cervical spinal plates Thoracolumbar and lumbosacral spinal and buttress plates Spine rod Interspinous plates Bone cement plug Centraliser Pin plug/port plug/blind plug Lead end cap (rubber) Blanking plugs for acetabular screw holes Non expandable cage Augments General (endosseous) implant Washer Nut Bolt Screw hole plug Screw-on sleeve Suture pledgets Multi component screws, such as:
Gastro intestinal clip Vascular anastomosis couplers Staple line reinforcement material | Locking rings Intramedullary nails Interspinous spacer Tympanostomy tubes Biliary stent |
Medical devices intended to be absorbed within 6 months
The following medical devices that are intended to be wholly, or mostly, absorbed by a patient’s body within 6 months of being implanted are exempted from the requirement to provide patient information materials. Examples of each type of medical device are provided below.
For use as a filler
Exempt |
---|
Viscosupplements for dermal or intra- articular applications |
Synthetic bone grafts |
Bone fillers |
For haemostasis
Exempt |
---|
Absorbable collagen based materials |
Haemostatic matrices |
Bonewax |
Collagen-based hemostatic vascular closure device |
For tissue approximation
Exempt |
---|
Tissue adhesives |
Absorbable tape for wound closure |
Absorbable staple line reinforcement material |
For fixation of other medical devices within tissue
Exempt |
---|
Absorbable antibacterial envelopes |
Similar medical devices
Further examples of Devices that are/are not exempt from requirements include:
- Exempt as these devices do not meet the definition of an implantable medical device:
- Eye irrigation solutions
- Opthalmic Viscoelastic Devices – used during surgery for 20-30 mins
- Distractors (non-implantable)
- Multi-Lead Trialing Cable for Spinal Cord Stimulation
- Tunneling tools.
- Not exempt as these devices meet the definition of an implantable medical device and are not exempted under Clause 13A.1(b) and 13A.1(2):
- Implantable tissues
- Absorbable stents.
Attachment 3: Best practice for patient information leaflets and patient implant cards
In addition to the legislated requirements, there are other features of leaflet and card design that can be very helpful for patients. The below information, which includes feedback from patients, is not mandatory, but is included to further improve the way this information can best be provided.
General design principles
When designing your leaflet or card, think about the recipient of the device by considering the:
- age of users
- target patient group
- literacy of users and
- visual acuity.
This part of the guidance will assist you with some of these considerations.
Use simple language
Wherever possible, plain language should be used so that information is easy to understand. Vague and unnecessarily complex language should be avoided. Manufacturers or sponsors may wish to use readability assessment programs available in many word processing programs.
Leaflets that are very long or unnecessarily complex may not be useful to patients (in addition to being unlikely to meet the requirement that they be readily understood by patients).
User-friendly design
You should consider the recipient of the device and any specific requirements they might have.
For example, if your device is likely to be implanted in the elderly, you may consider using larger text than the minimum requirement.
Use of images in the leaflet
It can be useful to use pictures or images (diagrams or drawings) to describe the device. For example, images showing where on the body the device would be implanted, or a list of where the device may be implanted could be helpful to patients.
If images are used, they must not be used in such a way as to promote a particular device or make or model of the device, over other alternative therapies or devices (see Compliance with advertising legislation).
Colour contrast
Colour contrast is an important tool in ensuring legibility of text for consumers and it may facilitate better understanding of the device and its functionality.
The Vision Australia colour contrast analyser can be used to assist you in deciding on how to present your text. This is available on the Vision Australia website.
Using other aspects in addition to colour
Individuals can perceive colours differently, some people are colour-blind and colours can look different in different lighting conditions. For these reasons, if colour was the only element used to distinguish information on a patient implant card for example, it may be difficult or confusing to identify the required information and the TGA may consider that the information in the leaflet is not written in a way that is readily understood by patients (see Clause 13A.3(4), Schedule 1, part 2 of the MD Regulations).
Attachment 4: Breast Implants - specific information to be included about adverse events
The following tables outline the known adverse events (side effects/complications) and potential adverse symptoms demonstrated against breast implant devices as at the time of publication of this guidance7. See Adverse events – urogynaecological meshes and breast implants in the main part of this guidance for further details.
No. | Known adverse events |
---|---|
1 | Capsular Contracture - occurs when the scar tissue or capsule that normally forms around the implant tightens and squeezes the implant, potentially causing hardness, pain or deformity |
2 | Breast pain, including extension to axillae and chest wall |
3 | Changes or loss in sensation to the nipple, breast, or skin over the breast |
4 | Rupture- intracapsular, extracapsular, or silent - of silicone filled implants |
5 | Deflation (+/- rupture) of saline filled implants |
6 | Asymmetry (one breast appears different in size or shape to the other) |
7 | Breast tissue thinning |
8 | Delayed wound healing |
9 | Skin breakdown and extrusion of the implant |
10 | Haematoma |
11 | Seroma |
12 | Infection and/or Inflammation |
13 | Malposition and/or Displacement |
14 | Ptosis |
15 | Skin rash |
16 | Wrinkling, folding or Rippling |
17 | Dissatisfaction with the result |
18 | Breast implant associated cancer |
19 | Lumps or collections in breast and axillae |
Attachment 5: Urogynaecological meshes – specific information to be included about adverse events
The following table outlines the known adverse events that may be associated with urogynaecological meshes at the time of publication of this guidance8. See Adverse events – urogynaecological meshes and breast implants in the main part of this guidance for further details.
No. | Known adverse events |
---|---|
1 | Punctures or lacerations of vessels, nerves, structures or organs, including the bladder, urethra or bowel. |
2 | Transitory local irritation at the wound site. |
3 | Foreign body response i.e. wound breakdown, extrusion, erosion, exposure, fistula formation and/or inflammation. |
4 | Mesh extrusion, exposure, or erosion into vagina or other structures or organs. |
5 | Mesh may potentiate an existing infection. |
6 | Over-correction (too much tension applied to the tape) may cause temporary or permanent lower urinary tract obstruction. |
7 | Acute and/or chronic pain. |
8 | Voiding dysfunction. |
9 | Pain during intercourse (dyspareunia). |
10 | Loss of sensation during intercourse (apareunia). |
11 | Pain or discomfort to the patient’s partner during intercourse (due to exposed mesh) |
12 | Neuromuscular problems including acute and/or chronic pain or weakness in the groin, thigh, leg, pelvic and/or abdominal area. |
13 | Recurrence of incontinence. |
14 | Bleeding including haemorrhage or haematoma. |
15 | Seroma. |
16 | De novo (new) or recurrent urinary incontinence. |
17 | Urinary frequency. |
18 | Urinary retention. |
19 | Adhesion formation. |
20 | Atypical vaginal discharge. |
21 | Mesh migration. |
22 | Allergic reaction/hypersensitivity. |
23 | Abscess. |
24 | Swelling around the wound site. |
25 | Recurrent prolapse. |
26 | Contracture. |
27 | Scarring. |
28 | Excessive contraction or shrinkage of the tissue surrounding the mesh. |
29 | Vaginal scarring, tightening and/or shortening (stenosis). |
30 | Constipation or defecation dysfunction. |
31 | Granulation tissue formation. |
32 | Wound breakdown (dehiscence) |
33 | Necrosis (tissue death) |
Glossary
Term | Definition |
---|---|
Active implantable device | See the definition in the Introduction. |
Implantable device | See the definition in the Introduction. |
Intended performance | Performance means the ability of the device to achieve its intended purpose as stated by the manufacturer. |
Intended purpose of the device | See definition in section 41BD(2) of the Act and Dictionary of the Therapeutic Goods (Medical Devices) Regulations 2002. |
Kind of medical device | See definition in section 41BE of the Act. |
Magnetic Resonance (MR) | Resonant absorption of electromagnetic energy by an ensemble of atomic nuclei situated in a magnetic field. |
Manufacturers of medical devices | See definition in section 41BG of the Act. |
Residual risks | This means any potential risks that remain that are associated with the use of the device that is outside the risks already identified and the precautionary steps identified by the manufacturer. |
Unique Product Identifier (UPI) | This means the unique product identifier given to the device by its manufacturer to identify the device and any variants. |
Printed or graphic information on the medical device or packaging | Printed information supplied on (or with) the device or packaging. Includes information identifying:
|
Footnotes
- Therapeutic Goods (Medical Devices) Amendment (Implantable Medical Devices) Regulations 2017, amending the Therapeutic Goods (Medical Devices) Regulations 2002.
- Where a Unique Product Identifier (UPI) is applicable to the device, the name of the device to be included in the Patient information leaflet must be the UPI. The UPI is applicable for devices listed under regulation 1.6 of the Therapeutic Goods (Medical Devices) Regulations 2002
- This is a reference to subclause 2(2) of the Essential Principles: see the MD Regulations.
Materials and substances included in the device, including (but not limited to) manufacturing residuals, medicinal substances, materials of microbial origin, stable derivatives of human blood or human plasma and animal origin materials that could potentially pose a risk to patients, are to be included in the leaflet (consumer device information). This is consistent with ensuring devices are safe, and aligns with Article 18(1)(d) of the EU MDR which requires the provision of ‘any other information to ensure safe use of the device by the patient, including the information in point (u) of Section 23.4 of Annex I.’
The TGA interprets Article 18(1)(d) to include not only materials intended to be in contact with a patient, but also any materials that may contact a patient through unintentional means, such as through manufacturing residues, leaking, leaching etc., as they could pose a potential health risk.
- The TGA website’s internal pages may require updating from time to time. The Patient Information Leaflet (PIL) should state the TGA website as https://www.tga.gov.au and direct the patient or consumer to report a problem or adverse event.
- This is a reference to the Global Medical Device Nomenclature System Code, as set out in ISO 15225:2000(E): see s 41BE(3) of the Therapeutic Goods Act 1989 and regulation 1.7 of the MD Regulations.
Supporting documents
Page history
Minor clarifications and amendments.
Amend relevant parts including Attachment 2: Implantable medical devices exempted from requirements to add additional devices and provide clarity.
Amend to include information on Regulation changes related to exempt items.
Amended to include information on obtaining consent to import, supply or export a medical device that does not meet the Essential Principles for sponsors that will not have compliant patient information materials by 1 December 2021.
Amend to include information on Regulation changes in relation to reduced consent application fees, exempt items and provision of electronic patient information materials.
Updated Attachment 2: Implantable medical devices exempted from requirements to add additional devices and provide clarity
Amendment to Attachment 1: Timetable for Transition to more clearly explain the operation of the transition timeframes
Re-order and make expression of some information clearer, add further detail about timing and presentation of PICs and PILs (including adverse events for certain devices), and consolidate other TGA website information on the same subject matter into this document
Update to e-leaflets and to correct the dates sponsors and manufacturers are to comply with implant cards or leaflets.
Original publication.
Minor clarifications and amendments.
Amend relevant parts including Attachment 2: Implantable medical devices exempted from requirements to add additional devices and provide clarity.
Amend to include information on Regulation changes related to exempt items.
Amended to include information on obtaining consent to import, supply or export a medical device that does not meet the Essential Principles for sponsors that will not have compliant patient information materials by 1 December 2021.
Amend to include information on Regulation changes in relation to reduced consent application fees, exempt items and provision of electronic patient information materials.
Updated Attachment 2: Implantable medical devices exempted from requirements to add additional devices and provide clarity
Amendment to Attachment 1: Timetable for Transition to more clearly explain the operation of the transition timeframes
Re-order and make expression of some information clearer, add further detail about timing and presentation of PICs and PILs (including adverse events for certain devices), and consolidate other TGA website information on the same subject matter into this document
Update to e-leaflets and to correct the dates sponsors and manufacturers are to comply with implant cards or leaflets.
Original publication.