The TGA is reviewing all ventilators, continuous positive airway pressure (CPAP), and bi-level positive airway pressure (BiPAP) devices in the Australian Register of Therapeutic Goods (ARTG) to determine whether the devices incorporate harmful sound abatement foam. This action follows the identification of risks posed by polyester-based polyurethane (PE-PUR) foam used as soundproofing material in devices supplied by Philips Electronics Australia Ltd.
The TGA has identified additional devices (other than those supplied by Philips Electronics Australia Ltd) that contain PE-PUR foam material for sound abatement purpose in the breathing gas pathway. The TGA has requested evidence regarding the long-term safety to ensure these devices are safe for use throughout their expected service life; some of this information is still under review by the TGA.
For sponsors who have not provided additional evidence, or provided timeframes for additional studies, the TGA has imposed conditions of inclusion (COI) on the ARTG entries to ensure the timely availability of this evidence to support their devices.
Below is the list of devices that contain PE-PUR foam material for sound abatement purposes. The TGA will update the list with current status as the review progresses. If hazards resulting in risks to health are identified for these devices, the TGA will undertake appropriate product safety notification or recall action.
Published: Information about Philips' recall action of correcting their affected devices.
ARTG entry | Sponsor | Manufacturer | Device type | Foam type | Models | Regulatory Actions/outcomes |
---|---|---|---|---|---|---|
140966 | Draeger Australia Pty Ltd | Draeger AG & Co KGaA | Ventilator, adult | PE-PUR | Carina Hospital Ventilator | Draeger has initiated recall of the devices already in market. Recall action is Product Defect Correction; further information can be located by searching the SARA database for 'Carina ventilator' at https://apps.tga.gov.au/PROD/SARA/arn-report.aspx - external site). |
152125 | Biomedical Solutions | F Stephan GmbH | Ventilator, neonatal/paediatric | PE-PUR | EVE NEO | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 152125' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a…). |
178669 | Medtronics Australia Pty Ltd | Covidien llc | Ventilator, home care | PE-PUR | Puritan Bennett 560 Ventilator - 409660 | Sponsor has provided additional data to support the long term safety. Submission is under review. |
178669 | Medtronics Australia Pty Ltd | Covidien llc | Ventilator, home care | PE-PUR | Puritan Bennett 560 Ventilator - 409660-06 | Sponsor has provided additional data to support the long term safety. Submission is under review. |
178669 | Medtronics Australia Pty Ltd | Covidien llc | Ventilator, home care | PE-PUR | Puritan Bennett 520 Ventilator - 4098300 | Sponsor has provided additional data to support the long term safety. Submission is under review. |
178669 | Medtronics Australia Pty Ltd | Covidien llc | Ventilator, home care | PE-PUR | Puritan Bennett 520 Ventilator - 4098300-06 | Sponsor has provided additional data to support the long term safety. Submission is under review. |
206250 | Boian Surgical Pty Ltd | Hoffrichter GmbH | Home CPAP unit | PE-PUR | Point 2 Auto CPAP | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data Further information can be located by searching the ARTG database for 'ARTG 206250' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a…). |
216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SMART | Sponsor has provided additional data to support the long term safety. Submission is under review. |
216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SOFTbsp; | Sponsor has provided additional data to support the long term safety. Submission is under review. |
216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SMART Plus | Sponsor has provided additional data to support the long term safety. Submission is under review. |